Frequently Asked Question
Data size is too big
Last Updated 10 years ago
Data size is a result of the following :
1.PQZip tolerance (compression efficiency)
2.Nominal setting
3.Stability of the network (noisy network create more data)
Make sure point 1+2 above are reasonable (for pqzip, the default settings are 0.1% for fix quality & 700MB for fix ratio)
Other suggestions:
1.If leakage channel not needed, disable it (this channel consume storage as any other channel and store up to 1024 samples per cycle)
2.Check reports are not enable and clean it's folder if needed.
If issue still present, please open a ticket and attach the following :
1.Configuration file (or pictures of Current, Voltage and recording settings).
2.Few pqzip files
3.Short description of the issue and the measurement.